1. Measuring range | Residual chlorine: 0-20.00mg/L, Resolution: 0.01mg/L; Temperature: 0- 99.9 ℃ Resolution: 0.1 ℃ |
2.Accuracy | better than ± 1% or ± 0.01mg /L |
3.Temperature | better than ± 0.5 ℃ (0 ~ 50.0 ℃) |
4.Minimum detection | 0.01mg /L |
5.Repeatability Chlorine | ± 0.01mg / L |
6.Stability Chlorine | ± 0.01 (mg / L)/24h |
7.Current isolated output | 4 ~ 20 mA(load <750 Ω) current output, measurement parameters can be selected independently (FAC, T) |
8.Output current error | ≤ ± 1% FS |
9. High and low alarm | AC220V, 5A, each channel can be selected independently measured parameters corresponding (FAC,T) |
10.Alarm hysteresis | can be set according to the selected parameters |
11.Communication | RS485 (optional) |
12.Working environment | Temperature 0 ~ 60 ℃, Relative humidity <85% It can be convenient to computer monitoring and communication |
13.Installation type | Opening type, panel mounted. |
14.Dimensions | 96 (L) × 96 (W) × 118 (D) mm; Hole Size: 92x92mm |
15.Weight | 0.5kg |
★ Application: Waste water, river water, swimming pool